User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Goldstar DM-9183 Digital Multimeter. DATE: 03-Jan-94 AUTHOR: User Contributed REVISION: 0.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 45 NUMBER OF LINES: 89 CONFIGURATION: Fluke 5440B CONFIGURATION: Fluke 5200A CONFIGURATION: Fluke 5215A CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 5220A CONFIGURATION: Fluke 5450A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DRAW 2 70 2 62 88 8 66 91 1.002 SET ON/OFF: ON [draw55,16,f] 1.003 DISP Let the unit warm up a few minutes. 1.003 DISP The input for Voltage is here, COM & V [234]. 1.003 DISP [draw11,75,f] 1.004 HEAD {Direct Volts.} 1.005 5440 200 190.00mV 0.5% 0.1U 2W 2.001 5440 200 -190.00mV 0.5% 0.1U 2W 3.001 5440 2 1.9000V 0.5% 0.001U 2W 4.001 5440 2 1.0000V 0.5% 0.001U 2W 5.001 5440 2 0.5000V 0.5% 0.001U 2W 6.001 5440 2 -1.9000V 0.5% 0.001U 2W 7.001 5440 20 19.000V 0.5% 0.01U 2W 8.001 5440 200 190.00V 0.5% 0.1U 2W 9.001 5440 1000 950.0V 0.5% 1U 2W 10.001 HEAD {Alternating Volts.} 10.002 5200 200 190.00mV 0.8% 0.3U 1kH 2W 11.001 5200 2 1.9000V 0.8% 0.003U 1kH 2W 12.001 5200 20 19.000V 0.8% 0.03U 1kH 2W 13.001 5200 200 190.00V 0.8% 0.3U 1kH 2W 14.001 5200 1000 950.0V 1.2% 3U 1kH 2W 15.001 HEAD {Direct Current.} 15.002 DISP The lower current ranges use the mA uA & COM input. 15.002 DISP [draw11,50,f] 15.003 5100 200 190.00uA 0.8% 0.1U 2W 16.001 5100 2 1.9000mA 0.8% 0.001U 2W 17.001 5100 20 19.000mA 0.8% 0.01U 2W 18.001 5100 200 190.00mA 0.8% 0.1U 2W 19.001 5100 2 1.9000A 8% 0.001U 2W 20.001 DISP The input for the 10 Amp range is A & COM. [draw11,25,f] 20.002 5220 10 9.500A 2% 0.1U 2W 21.001 HEAD {Alternating Current.} 21.002 5220 10 9.500A 3% 0.1U 1kH 2W 22.001 DISP Move back to the lower range current input, mA uA & COM. 22.001 DISP [draw11,50,f] 22.002 5100 2 1.9000A 1% 0.003U 1kH 2W 23.001 5100 200 190.00mA 1% 0.3U 1kH 2W 24.001 5100 20 19.000mA 1% 0.03U 1kH 2W 25.001 5100 2 1.9000mA 1% 0.003U 1kH 2W 26.001 5100 200 190.00uA 1% 0.3U 1kH 2W 27.001 HEAD {Resistance.} 27.002 DISP Connect back to the COM & V [234] input. [draw11,75,f] 27.003 5450 200 1.00Z 0.5% 0.3U 2W 28.001 5450 200 1.90Z 0.5% 0.3U 2W 29.001 5450 200 10.00Z 0.5% 0.3U 2W 30.001 5450 200 19.00Z 0.5% 0.3U 2W 31.001 5450 200 100.00Z 0.5% 0.3U 2W 32.001 5450 200 190.00Z 0.5% 0.3U 2W 33.001 5450 2 1.0000kZ 0.5% 0.001U 2W 34.001 5450 2 1.9000kZ 0.5% 0.001U 2W 35.001 5450 20 10.000kZ 0.5% 0.01U 2W 36.001 5450 20 19.000kZ 0.5% 0.01U 2W 37.001 5450 200 100.00kZ 0.5% 0.1U 2W 38.001 5450 200 190.00kZ 0.5% 0.1U 2W 39.001 5450 2 1.0000MZ 0.5% 0.001U 2W 40.001 5450 2 1.9000MZ 0.5% 0.001U 2W 41.001 5450 20 10.000MZ 1% 0.02U 2W 42.001 5450 20 19.000MZ 1% 0.02U 2W 43.001 HEAD {Continuity Tester.} 43.002 SET Func. Sw.: Cont. ))) [draw44,50,f] 43.003 5450 100Z S 2W 43.004 EVAL {Does the unit indicate about 100 Ohms AND does it sound?} 44.001 5450 190Z S 2W 44.002 EVAL {Is the unit silent?} 45.001 SET Func. Sw.: 1KV DC [draw44,50,f] 45.002 SET ON/OFF: OFF [draw55,16,f] 45.003 END